Legal metrology

Calibration, Verification and Re-verification of Marine Scales

Updated 5 min readBy WPL Industries Engineering
Short answer

Calibration measures how far a scale's indication deviates from reference weights, with an uncertainty. Verification is a legal conformity check that ends with a mark or certificate. Re-verification intervals are set nationally, not by EU law. For fisheries weighing, scales must be approved, calibrated and sealed under national law, and records must be available to inspectors.

Calibration, adjustment, verification and re-verification are often used as if they meant the same thing. They do not, and on a fishing vessel the difference decides whether a scale may be used for catch records. This article defines each term using the international vocabularies, explains test weights, shows how re-verification intervals are set nationally and gives a practical routine for checks at sea. This article is general information, not legal advice: confirm the requirements for your vessel with your national fisheries control authority and metrology authority.

Calibration, adjustment and verification defined

Term Definition Result Source
Calibration Establishes the relation between values provided by measurement standards and the instrument's indications, with measurement uncertainties Calibration certificate with errors and uncertainty; the scale itself is not changed VIM 2.39
Adjustment Changing the instrument so that it indicates correctly; VIM warns not to confuse it with calibration Changed span or zero; seals may need to be broken VIM 2.39, note 2
Verification Conformity assessment (other than type evaluation) that results in a verification mark or certificate Pass or fail against legal limits; mark, seal or certificate VIML 2.09
Initial verification Verification of an instrument that has not been verified before First legal mark VIML 2.12
Subsequent verification Any verification after a previous one: mandatory periodic, after repair, or voluntary Renewed validity VIML 2.13
Mandatory periodic verification Subsequent verification at intervals laid down by regulations What is usually called re-verification VIML 2.14

In short: calibration tells you how wrong the scale is; verification tells you whether that is legally acceptable. Sources: JCGM 200:2012 (VIM) and OIML V 1:2013 (VIML).

Who does what over the life of a scale

  1. Before first use: the manufacturer, with a notified body where the module requires it, completes conformity assessment and affixes CE and M marking for regulated uses (see NAWI and MID directives explained).
  2. Approval for fisheries weighing: systems used under Article 60 of the Control Regulation must be "approved, calibrated and sealed in accordance with national law"; for on-board weighing the flag Member State ensures this (Article 5(1)-(2) of Implementing Regulation (EU) 2026/1932, applicable from 11 January 2027). Until 10 January 2027, Article 72 of Regulation 404/2011 still requires systems to be calibrated and sealed under national systems, with a calibration record kept by the person responsible.
  3. In service: periodic verification, verification after repair and market surveillance are national matters; WELMEC Guide 2:2023 states that in-service control and periodic verification of NAWIs installed in ships "shall be dealt with nationally".
  4. Every day: the operator carries out routine checks and keeps records.

Test weights: which class is good enough?

Weights used to verify a scale should meet OIML R 111 and have an error no greater than one third of the scale's maximum permissible error for the applied load (OIML R 76-1, clause 3.7.1). The maximum permissible errors of common weight classes are (R 111-1, Table 1):

Nominal value F1 F2 M1 M2 M3
5 kg ±25 mg ±80 mg ±250 mg ±800 mg ±2.5 g
10 kg ±50 mg ±160 mg ±500 mg ±1.6 g ±5 g
20 kg ±100 mg ±300 mg ±1 g ±3 g ±10 g

Worked example. A class III scale with a verification scale interval of 10 g, loaded with 20 kg (2 000 verification scale intervals), has a maximum permissible error on verification of ±10 g (NAWI Directive Annex I, Table 3). One third is 3.3 g, so a 20 kg M1 or M2 weight qualifies, but an M3 weight (±10 g) does not. For a scale with a verification scale interval of 1 g loaded with 5 kg (5 000 intervals), the limit is ±1.5 g; one third is 0.5 g, so a 5 kg M1 weight qualifies and M2 does not.

WELMEC Guide 2 notes that EN 45501 requires a NAWI to be verified by means of weights, and that for vehicle-mounted instruments it can be difficult to place weights; the same practical problem applies on deck, so plan how weights will be handled safely.

Re-verification intervals are set nationally

EU law does not set a single re-verification interval. Germany illustrates how detailed national rules can be: the default verification period is two years (MessEV section 34), while Annex 7 sets three years for non-automatic scales with a maximum capacity of 3 000 kg or more, other than scales for building materials (item 2.2.1), and one year for automatic checkweighers including automatic grading scales (2.3.1) and for automatic scales with label printers used to make up prepackages of unequal quantity (2.3.2). Other Member States use different periods or none at all for some categories, so ask the authority of the country where the scale is used.

Verification after repair is a separate trigger (VIML 2.13). Breaking a seal to adjust a scale can end the validity of a verification under national rules, so plan adjustments with the verifier.

Records to keep

Record Why Legal reference
Weighing record per landing Kept for three years by the weighing operator Reg. 1224/2009 Art. 60(5)
System details, latest calibration certificate with expiry date, seal numbers, cumulative-weight data, wiring schematics Must be provided to officials during weighing inspections Delegated Reg. (EU) 2025/1766 Art. 32(1)(a)
On-board weighing authorisation and certification documents of the on-board systems Kept on board or digitally accessible Implementing Reg. (EU) 2025/2196 Art. 8(4)
Log of all use of conveyor belt systems, including maintenance, approval, reruns and washing Minimum requirement for belt systems Reg. 2026/1932 Annex I, point 2(e)
Failures of belt, batch or weighbridge systems Notify within 24 hours Reg. 2026/1932 Art. 5(3)(b)
Daily check results, repairs and parameter changes Evidence of control between verifications Good practice

Article 32 of Delegated Regulation (EU) 2025/1766 has applied since 10 January 2026. What a calibration certificate for a non-automatic scale should contain, and how to read its uncertainty, is explained in EURAMET cg-18 calibration explained. The Article 60 rules themselves are summarised in EU weighing rules for catches.

Practical routine checks on board

Routine checks do not replace legal verification, but they catch problems before an inspector does. A workable routine at the start of each watch or landing:

  1. Clean and inspect. Remove fish residue, ice and standing water from the platform and around the load receptor; debris touching the frame causes errors. See cleaning and washdown of scales.
  2. Check seals and cables. Confirm seals are intact and cables and glands undamaged.
  3. Zero. With the platform empty, zero the scale and confirm a stable zero indication.
  4. Tare. Place an empty box of the type in use and confirm the tare value; Regulation 2026/1932 requires zeroing per box, container and pallet or pre-programmed manufacturer's tare (Annex I, point 1(c)).
  5. Check weight. Place a test weight near the typical load, preferably in calm conditions, and record the indication. Repeat three times, and at the centre and corners for platform scales.
  6. Compare and act. Compare with an internal action limit that is tighter than the in-service maximum permissible error, which is twice the verification limit (NAWI Annex I, point 4.2). If the limit is exceeded, stop using the scale for regulated weighing and switch to another approved system (Article 5(3), Regulation 2026/1932).
  7. Record date, time, results, sea state and the name of the person checking.

Repeat the check after heavy weather, impacts, repairs or relocation. How motion affects readings is explained in why ordinary scales fail at sea.

How WPL approaches this

The R10 Configuration Panel supports calibration, maintenance, diagnostics and backup and recovery of scale data; advanced settings are intended for trained personnel only. WeightControl logs every weighing automatically; see weighing data logging. More on the regulatory framework is on the legal metrology hub.

Frequently asked questions

Is a calibration certificate enough to use a scale for catch records?

Usually not on its own. A calibration certificate documents errors and uncertainty, but weighing under the Control Regulation requires systems approved, calibrated and sealed in accordance with national law. Many countries also require legal verification with a verification mark. Ask your fisheries control authority which documents it accepts for your vessel.

Can the crew adjust the scale span themselves?

Crew can zero and tare, which are normal operating functions. Adjusting span or other metrological settings normally means opening sealed parameters, which can end the validity of a verification and trigger verification after repair under national rules. Leave adjustment to authorised service staff and record every intervention.

How often should test weights themselves be calibrated?

OIML R 111 defines weight classes and their maximum permissible errors but does not set one interval for every user. Base the interval on use, handling and storage conditions, and the requirements of your verifier or accredited laboratory. Weights stored in a wet, salty environment or dropped on deck should be checked sooner.

Why check a scale at the centre and the corners?

Weighing results must be sufficiently insensitive to changes in the position of the load on the load receptor, one of the essential requirements in Annex I of the NAWI Directive. A corner check with the same weight reveals a damaged load cell, debris under a corner or a platform touching its frame, which a centre-only check can miss.

Sources

  1. JCGM 200:2012 International vocabulary of metrology (VIM), 3rd edition
  2. OIML V 1:2013 International vocabulary of terms in legal metrology (VIML)
  3. OIML R 111-1:2004 Weights of classes E1, E2, F1, F2, M1, M1-2, M2, M2-3 and M3
  4. OIML R 76-1:2006 Non-automatic weighing instruments, Part 1: Metrological and technical requirements - Tests
  5. Directive 2014/31/EU on non-automatic weighing instruments (NAWI Directive)
  6. WELMEC Guide 2:2023 Guide for NAWI, AWI and MDMI (Directives 2014/31/EU and 2014/32/EU), section 2.3.1 NAWIs installed in ships
  7. German Mess- und Eichverordnung (MessEV), section 34 Eichfrist
  8. German Mess- und Eichverordnung (MessEV), Anlage 7 (verification periods)
  9. Commission Implementing Regulation (EU) 2026/1932 on the weighing of fishery products (OJ L, 11.8.2026)
  10. Commission Implementing Regulation (EU) No 404/2011 (weighing rules, Articles 69-77)
  11. Commission Delegated Regulation (EU) 2025/1766 on control, surveillance, inspection and enforcement (Article 32)
  12. Commission Implementing Regulation (EU) 2025/2196 laying down detailed rules for Regulation (EC) No 1224/2009 and repealing Implementing Regulation (EU) No 404/2011

Written and reviewed by WPL Industries weighing engineers. Technical and regulatory content is checked against the cited sources. Editorial policy